Cortical Bone Plate Properties Assessment using Inversion of Axially Transmitted Low Frequency Ultrasonic Guided Waves
This study investigates the use of low-frequency ultrasonic guided waves to evaluate bone quality and mechanical properties, with potential applications in early osteoporosis detection. A specialized ultrasonic transducer operating below 500 kHz was used to collect data from bone phantom plates. The researchers extracted experimental dispersion curves using 2D Fourier analysis and compared them with simulated curves generated using a semi-analytical isogeometric analysis (SAIGA) model. The estimated bone mechanical properties differed by less than 5% from independent bulk-wave measurements, demonstrating the potential of guided wave ultrasound as a non-invasive method for assessing cortical bone properties.